  1. Survey all diagnostic X-ray equipment in accordance with the National Council on Radiation Protection and Measurements (NCRP), Joint Commission on Accreditation of Health Organization (JCAHO) requirements and various regulatory agencies.  The survey is to include the following performance evaluations, safety evaluations, and output measurements:

     

    • Light beam to X-ray beam alignment
    • Accuracy of beam size indicators
    • Overall performance of radiographic fluoroscopic collimators
    • Output versus kVp, mAs and distance
    • Proper function of interlocks, timers and exposure switches

     

  2. Perform tube performance evaluation on each radiographic unit which includes survey as outline above plus:

     

    • Radiographic time accuracy
    • kVp accuracy
    • Reproductibility of kVp, timer and output
    • MA linearity
    • Accuracy of distance indicators
    • Half-value layer determination and appropriateness of total filtration
    • Evaluation of Automatic Exposure Control balance and performance

     


  3. Perform tube performance evaluation of each fluoroscopic unit which includes as outlined in the first paragraph, plus:

    • kVp accuracy
    • Exposure rate vs kVp
    • Exposure rate in Automatic Exposure Control through use of appropriate phantoms
    • Low contrast resolution of the imaging system
    • High contrast resolution of the imaging system
    • Half-value layer determination and appropriateness of total filtration 
    • Evaluation of output in other imaging modes

     


  4. In addition, for all radiographic equipment:

     

    • Monitor patient exposures for techniques and equipment.  Provide a summary of measured exposures with comparable national standards in accordance with JCAHO standard DR.2.2.4.2
 